Shopping Cart     Empty        
    Welcome, sign in    Private person / Company / Distributor
Monitored Defibrillator. Biphasic truncated exponential  
Photo gallery
Reference Availability Amount Purchase
207-CUHD1   
6,495.00€
7,014.60€ (VAT Incl.)
207-CUHD1-SPO2 With pulse oximetry probe    32-42 D
7,495.00€
8,094.60€ (VAT Incl.)
:::
Options

207-CUHD1: With printer, and cable 3 leads ECG
207-CUHDIV1: With printer cable 3 leads ECG, SpO2, and pacemakers.
207-CUHDIV2: With printer cable, 12-Lead ECG, SpO2, and pacemakers.
207-CUHDIV3: With printer cable, 12-Lead ECG, SpO2, NIBP, and pacemakers.

Defibrillation biphasic truncated exponential waveform. e-cube
Monitoring NIBP, SpO2, ECG (12 Dervan.) Memory and trends.
External pacemaker.
DEA integrated.
Built-in printer.

ECG MONITORING

Patient connection: Wheel or disposable defibrillation electrodes
Cable 3, 5, or 10 leads with ECG electrodes.
Deriv. monitored: I, II, III, aVR, aVF, aVL, V1, V2, V3, V4, V5, V6
Bandwidth: 0.3 ~ 40Hz (-3dB).
ECG size: 5, 10, 20 mm / mV - display GráficoECG.
Auto scaling of ECG wave.
Heart Rate: Digital 30 ~ 300 bpm. (± 3bpm).
Alarms: Configurable FC, minimum and maximum asystole, VF, VT.

SpO2 pulse oximetry (Nellcor Oximax)

Saturation: 70-100% (± 3 digits)
Pulse Ratio: 20 to 250 bpm. (± 3 digits)
Infusion: 0.2%. Configurable alarms.

DISPLAY

Display Type: LCD high-resolution graphic
Screen Size: 7 "(152mm x 91mm)
Resolution max. : 800 x 480 pixels.
Sweep speed: 25mm/seg. nominal
Display Time: 3.2 sec. (ECG)
Information: Energy discharge, Synchronization, heart rate,
Palas Connectors / pads, Alarms. On / Off, Function and status AED, Defibrillator function test, correction of mistakes and errors. Curve Curve ECG and SpO2 plethysmograph (optional)

DEFIBRILLATOR
Operational Mode: - Semi DEA AAMI DF-80, ERC 2005 - Manual. By PADS or disposable blades.Deriv selection: I, II, III, aVR, aVL, aVF, V. - Wheel / Pads. ECG ext.
Waveform: Biphasic truncated exponential e-cube
Power: Semi-Automatic: 150 J 50 ohm load. (Preset)
Variable energy. Selection of levels.
(150J, 150, 180J / 150J, 180J, 180J)
(50J, 50J, 100J / 50J, 100J, 100J)
Self-adjusting power according to patient impedance.
Charging time: less than 10 seconds.
Manual: Selecting energy variable. (20 levels)
1-10j, 15j, 20J, 30J, 50J, 70J, 100J, 120J, 150J, 170j, 200j.
Internal paddles: 1-10j, 15j, 20J, 30J, 50J.
Patient Impedance: 25 Ohm to 175 Ohm, unbalanced.
Charge time: AC Adapter: less than 5 seconds to 200J.
Battery: less than 7 seconds to 200J
Sync Cardioversion: Energy Transfer for 60 ms. QRS peak.
Sensitivity and specificity 99.5%
Results according to AHA / IMT ECG Database.
Detection level> 0.10 mV ECG
Charging time: less than 10 seconds. (AED mode)
Pads:
Multi-functional electrodes. Disposable.
Adhesive pads for adults.
CONTROLS
On / Off, Download (Shock) Buttons. Energy selector. Cardiosin-sync, bud analysis / Load Button Pause / downloads.Volume and screen menu controls. Print button.
PACEMAKER
Curve: monophasic truncated exponentialActivated by the user
Amplitude: 0 ~ 200 mA (± 5mA)
Pulse width: 20 ms (± 1.5%)
Heart Rate: 30 ~ 180 ppm (± 1.5%)
Refractory Period: 340 msec (30 to 80ppm); 240mseg (90 to 180ppm)
NOTICES OF VOICE AND MESSAGE DISPLAY

Audible in Castilian, which guide the user through the ERC protocol 2005 in DEA. Example: Connect electrodes - Electrode Check - Do not touch the patient - Download - Check patient. etc.
Screen prompts are easy to interpret that guide the user through the proper protocol for action.

MEMORY CARD DATA and PC
ECG display, ability to see historical ECGEvents 6 hours recording ECG curves and 1 / 2 hour of recorded events or voices (depending on the SD card used)
Taxable LINEUP: Time, Date, FC, Downloads realized, unrealized downloads, download recommendations, or not, ECG Analysis, Test daily, weekly and monthly test and manuals, on-screen ECG gating.
Automatic self test
Internal electronic test, analysis of the receiving system of curves and battery capacity.Check daily, weekly and monthly data stored. of environmental operating conditions and equipment.
EXTERNAL CONNECTIONS
Infrared port RS232 UART / IrDA.Transmitted data: ECG events, system software, test results, transmission between the unit and a PC.
Power adaptor & battery
Internal Rechargeable Battery. / 12V 4.2 Ah Ni / MH. 200 Downloads. (200j). Recharge time: Full charge in 4 hours.External Battery Pack (Pack of safety). Disposable. 14.8 V 3.1 Ah LiMnO2. 100 Downloads. (200j).
AC / DC Adapter Input :100-240V 50/60Hz 170 VA. Output: DC 18V, 5A
SPECIFICATIONS
Galvanic Protection: BF TypeEnvironment:
- Operating: Temp. 0 º C to 40 º C
Humidity: 5% to 95%
- Storage and transportation:
Temperature. -20 º C to 60 º C
Dimensions:

Size: 318 x 208 x 355 mm
Weight: 4.7 kg with shovels.



PERHAPS YOU WILL ALSO BE INTERESTED IN


Continue shopping